Quantitative NMR by the ERETIC method
We have implemented a variation of the ERETIC method designated
"digital ERETIC" (S. Akoka and M. Trierweiler, "Improvement of the
ERETIC Method by Digital Synthesis of the Signal and Addition of a
Broadband Antenna Inside the NMR Probe," Instrumentation Science & Technology, 30, 21-29 (2002)). In this
version of the experiment, a constant amplitude synthetic signal is
added to the experimental spectrum in the form of an exponentially
decaying shaped proton pulse routed to the carbon coil of the NMR
probe. The synthetic peak will be at -1 ppm (it can be moved if
you have a real signal there, although that will seldom be the case).
The real signals can be integrated relative to this synthetic
signal. It can be used just like adding a known quantity of some
reference material to the NMR solution, except nothing actually needs
to be added to, or later removed from, a pure sample.
You can rpar the parameter set ERETIC and be ready to do the
experiment. As set up, the parameters call for a single scan with
no dummy scans. These conditions were chosen to avoid issues with
differential relaxation of different protons in the analyte--the first
scan is always quantitative. Unlike most other experiments we do,
it is best not to execute the rga command to scale the receiver gain,
because the sample is pulsed repeatedly during that procedure. If
you do an rga, wait a couple of minutes before acquiring the data you
intend to use. If you wish to take mutiple scans in the spectrum,
change ns to the desired number of scans and I recommend a d1 of at
least 60 seconds in that case. Both the default receiver gain and the
size of the
synthetic signal are chosen to match a typical multi-milligram
sample. The signal can be scaled up or down in absolute size by
changing the power level, pl14, of the shaped pulse. The default pl14
is 37 dB and a change of 6 dB units will change its size by a factor of
2 (increasing the number makes the peak smaller). More generally, the
change in intensity can be calculated by dividing the change in dB by
20, and then taking the inverse log of that number (e.g. invlog (6/20)
= factor of 2). Don't use a pl14 setting less than 30 in any case.
The only other change you might need to make would be to the carrier
position, o2, of the synthetic signal. The real signals in the
spectrum will have phase corrections that depend on the electrical
tuning of the probe, which can vary from solvent to solvent.
These changes do not affect the phase of the synthetic signal, but that
signal's phase is very dependent on the precise frequency offset
between it and the spectrometer carrier frequency o1. You can set
o2 to the nearest 0.01 Hz, and a 0.01 Hz offset will change the phase
of the line by about 1 degree; phase correct the real peaks and then
adjust o2 to get the best match between the phases of the synthetic
and real signals in response to day to day changes in probe
tuning. That can typically be done by varying o2 by 0.01-0.02 Hz
(if
needed).
To determine the concentration of your sample, phase the spectrum,
baseline correct it with abs, then set the integral of the synthetic
peak to 21.8. If you then look at the integral of a peak in the
analyte which should count for 1 proton, its intensity will give the
analyte concentration in millimolar. Multiply that value by the
solution volume in ml to obtain the number of micromoles present;
multiplying that by the formula weight will give an absolute weight in
micrograms. If pl14 does not have the default value of 37, adjust
the assigned integral accordingly--for example, if you lower pl14 from
37 to 43, the 6 dB change will reduce the size of the synthetic signal
by half. In that case, assign its integral to be 10.4.
ERETIC is available as an automation experiment in ICON-NMR with the
user-adjustable parameters of d1, ns, and o2. It would be
advisable to run the first sample manually and note any non-default
settings, particularly for o2, and then use the experimentally
determined values during the automation run. You can add any
other automation experiments that you wish to run on the samples at the
same time.
NMR integrals are subject to a few percent error owing to noise in the
spectrum, shimming problems giving rise to nonideal lineshapes, and so
forth. Different NMR tubes may also have slightly different inner
diameters which will lead to different active volumes and a resultant
systematic error in the calculated concentration. If no effort is
made to control these factors,
and you don't do multiple runs in order to get statistics, it is
probably reasonable to assign 10% accuracy to the ERETIC-derived
value. But it is quite fast and easy to do--certainly a lot
faster than drying the sample down and weighing the solid.
The first place the experiment is implemented is on the Bruker 400 at
LSRL. This system works really well for digital ERETIC because
the carbon channel
and the lock channel are not on the same coil in the probe on that
instrument, and we have a 24 position NMR CASE sample changer to
automate the experiment there. We have tested ERETIC on the 400
in SBC, but in the probe there, the carbon and lock channel share a
coil. In this case the lock is severely disrupted during an
ERETIC experiment. We found it is still possible to get an ERETIC
spectrum, provided the lock and sweep are turned off on the BSMS
control panel prior to acquiring the data. This instrument does
not have automation in any case but it may still be useful to use
ERETIC there to spot check single samples.
Dave 9/7/06